Health product recall

MOSAIQ (2016-07-15)

Starting date:
July 15, 2016
Posting date:
July 29, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59596

Affected products

A. MOSAIQ

Reason

Retention of outdated information in an order set. It is possible that a change to an order set will not be saved in the current open care plan when the following work flow is utilized:

  1. open either the care plan edit form or the care plan instance form and leave the form open.
  2. select an order set that is part of the same care plan and make a change to the order set. This can be changed to the order set name and label or a change to the drug, dose, administration instructions, etc.
  3. the update made to the order set will not be saved in the current open care plan.

Affected products

A. MOSAIQ

Lot or serial number

VERSION 2.50.05 AND HIGHER

Model or catalog number

MOSAIQ

Companies
Manufacturer

Elekta Business Area Software Systems IMPAC Medical Systems, Inc.

100 Mathilda Place, 5th floor

Sunnyvale

94086

California

UNITED STATES