This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
MOSAIQ (2015-07-01)
- Starting date:
- July 1, 2015
- Posting date:
- July 8, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-54130
Recalled Product
- MOSAIQ
Reason
Incorrect field size for Stereotactic Cones. A problem exists in MOSAIQ resulting in the incorrect field size being sent to the treatment machine for Stereotactic plans using cones.
Affected products
A.MOSAIQ
Lot or serial number
VERSION: 2.41 AND HIGHER
Model or catalog number
MOSAIQ
Companies
- Manufacturer
-
Elekta Business Area Software Systems IMPAC Medical Systems, Inc.
100 Mathilda Place, 5th floor
Sunnyvale
94086
California
UNITED STATES