MOSAIQ
Brand(s)
Last updated
Summary
Product
MOSAIQ
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| MOSAIQ | version 3.1.3.3 | VERSION 3.10 |
Issue
MOSAIQ 2D image review application may provide incorrect image offsets when a 2D verification image is taken at a patient support angle (couch rotation) other than 0 degrees, and if the treatment machine is characterized at a scale that is not IEC61217. If the incorrect shift was not detected and used for treatment, mistreatment could occur.
Recall start date: January 15, 2025
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Elekta Solutions Ab
Hagaplan 4, Stockholm, Sweden, 11368
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76878
Get notified
Receive emails about new and updated recall and safety alerts.