Health product recall

MOSAIQ

Last updated

Summary

Product
MOSAIQ
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

MOSAIQ

version 3.1.3.3

VERSION 3.10

Issue

MOSAIQ 2D image review application may provide incorrect image offsets when a 2D verification image is taken at a patient support angle (couch rotation) other than 0 degrees, and if the treatment machine is characterized at a scale that is not IEC61217. If the incorrect shift was not detected and used for treatment, mistreatment could occur. 

Recall start date: January 15, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Elekta Solutions Ab

Hagaplan 4, Stockholm, Sweden, 11368

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-76878

Get notified

Receive emails about new and updated recall and safety alerts.