Mosaiq
- Starting date:
- June 30, 2017
- Posting date:
- July 18, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-64008
Reason
A table shift in the vertical direction was inverted during treatment. An inverse shift will result in the patient being out of position by double the shift magnitude. A mac file is instrumental to the correct operation, if values are incorrectly input, inverse shifts may occur.
Affected products
Mosaiq
Lot or serial number
Not Applicable
Model or catalog number
MOSAIQ
Companies
- Manufacturer
-
Elekta Business Area Software Systems IMPAC Medical Systems, Inc.
100 Mathilda Place, 5th floor
Sunnyvale
94086
California
UNITED STATES