Health product recall

Monoject Standard Hypodermic Needles (2020-09-28)



Starting date:
September 28, 2020
Posting date:
October 23, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74151


Last updated:

Affected Products

Monoject Standard Hypodermic Needles

Reason

This recall is being conducted due to a potential for the needle to detach from the cartridge after the bottom of the cartridge is twisted and removed. If the needle is not securely seated, the needle can fall out of the sheath. In some instances, the needle may be projected out of the cartridge, which could result in a needle stick/puncture from the clean/unused needle.

Affected products

Monoject Standard Hypodermic Needles

Lot or serial number

11144

Model or catalog number

8881250248

Companies
Manufacturer

Cardinal Health 200, LLC (carrying on the patient recovery business from Covidien LLC)

15 Hampshire Street

Mansfield

02048

Massachusetts

UNITED STATES