Monnal™ T60 Ventilator
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Monnal™ T60 Ventilator |
Not applicable. |
KA017115 |
Expiratory Valve Monnal EVA V2 |
Not applicable. |
KY732991 |
Issue
The Monnal EVA V2 (2nd generation - expiratory valve) used on Monnal T60/T60 Advanced ventilators can lead to under-delivery of volumes in "child/infant" patients due to poor membrane sealing. No patient incidents have been reported; the detection occurred during tests. Air liquide medical systems is recalling these valves to replace them with the first generation to prevent the risk of hypoventilation.
Additional information
Details
Air Liquide Medical Systems
6 Rue Georges Besse, Antony, France, 92182