Health product recall

Monnal T60 (2019-03-31)

Starting date:
March 31, 2019
Posting date:
April 29, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69714


Last updated:

Affected Products

Monnal T60

Reason

The internal battery, regarding the conditions and frequency of the solicitations, can have lost a significant capacity since 2 years of installation.  This significant capacity loss does not systematically allow the algorithm to flag the end of automonmy alarms (''low batteries'' and "batteries empty'').

Affected products

Monnal T60

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

KA010000

Companies
Manufacturer

Air Liquide Medical Systems

6 Rue Georges Besse, Parc de Haute Technologie

Antony

92182

FRANCE