Health product recall

Monnal T60

Dernière mise à jour

Summary

Produit
Monnal T60
Problème
Medical devices - Performance issue
Ce qu’il faut faire

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

Monnal T60

All lots.

KA017115
KA010000

Problème

During a ventilation sequence, the user started the "100% O2" function. This function lasts for two minutes but can be interrupted by pressing the 100% O2 key. Note: after the 100% O2 function has been used, the FiO2 setpoint originally entered is reapplied. The user wanted to change the FiO2 setpoint a few seconds before the end of the two-minute period of the 100% O2 function, and the confirmation of this setpoint took place just after the end of the 100% O2 function, which can cause a conflict between the setpoint applied and its display.

Recall start date: March 13, 2023

Additional information

Details
Original published date: 2023-03-24
Type d’avis ou de rappel
Health product recall
Catégorie
Health product - Medical device - Anaesthesiology
Entreprises

Air Liquide Medical Systems

6 Rue Georges Besse, Antony, France, 92182

Published by
Health Canada
Audience
Healthcare
Industry
Classe de rappel
Type II
Identification number
RA-73394