Health product recall

Monnal T60

Last updated

Summary

Product
Monnal T60
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalog number

Monnal T60

All lots.

KA017115
KA010000

Issue

During a ventilation sequence, the user started the "100% O2" function. This function lasts for two minutes but can be interrupted by pressing the 100% O2 key. Note: after the 100% O2 function has been used, the FiO2 setpoint originally entered is reapplied. The user wanted to change the FiO2 setpoint a few seconds before the end of the two-minute period of the 100% O2 function, and the confirmation of this setpoint took place just after the end of the 100% O2 function, which can cause a conflict between the setpoint applied and its display.

Recall start date: March 13, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies

Air Liquide Medical Systems

6 Rue Georges Besse, Antony, France, 92182

Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-73394

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe