Health product recall

MONACO (2021-09-14)



Starting date:
September 14, 2021
Posting date:
October 1, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76553


Last updated:

Affected Products

MONACO

Reason

Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

Affected products

MONACO

Lot or serial number
  • v 5.40.00
  • v 5.40.01
  • v 5.40.02
  • v 5.51.10
Model or catalog number
  • MONACO
Companies
Manufacturer

ELEKTA, INC.

1450 BEALE STRET, SUITE 205

St Charles

63303

Missouri

UNITED STATES