Health product recall

MONACO (2019-02-08)

Starting date:
February 8, 2019
Posting date:
August 2, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70599


Last updated:

Affected Products

MONACO

Reason

If "Improve Target Dose" was chosen as an optimization model in a previous treatment session, Monaco will automatically use this optimization model again when proceeding with the online plan adaptation of a completion plan when it should not.

Affected products

MONACO

Lot or serial number

12274-CMS1/5.40.0

12473-CMS1/5.40.0

Model or catalog number

MONACO

Companies
Manufacturer

Elekta INC.

13723 RIVERPORT DRIVE,, SUITE 100

Maryland Heights

63043

Missouri

UNITED STATES