Health product recall

MOLLI 2 Systems

Last updated

Summary

Product
MOLLI 2 Systems
Issue
Medical devices - Labelling and packaging
What to do

Contact the manufacturer if you require additional information.

Affected products

Affected products

Lot or serial number

Model or catalogue number

MOLLI OncoPen

Various

MP2015A

MOLLI 2 Tablet

Various

MT2001

MOLLI 2 Wand

Various

MP2001

MOLLI Introducer And MOLLI Marker

Various

MS-00009

MOLLI Introducer And MOLLI Marker

Various

MM1000

Issue

Stryker has updated the MOLLI 2 system Instructions for Use (IFU) following two medical device reports (MDRs) submitted to the FDA (June 5 2025). These reports involved patient injuries associated with MOLLI Marker dislodgement or dislocation due to magnetic attraction from magnetized surgical tools. The updated IFU now includes an additional warning that using magnetized surgical tools in close proximity to the marker during localization may lead to marker dislodgement.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Health products - Medical devices - General and plastic surgery
Companies
Stryker Endoscopy, A Div. Of Stryker Corp. Dba Stryker Endoscopy
5900 Optical Court, San Jose, California, United States, 95138
Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Recall date
Identification number
RA-81524

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe