Modular Cathcart Fracture Head
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
---|---|---|
Modular Cathcart Fracture Head Hip Ball (Endoprosthesis) |
All lots. |
1363-43-000 |
Modular Cathcart Fracture Head Tapered Spacer Articul-Eze Taper (Endopros.) |
All lots. |
1363-14-000 |
Issue
During an internal review, it was determined that the Cathcart head includes a +5mm offset, which was not specified in previous versions of the self-centering bipolar and Modular Cathcart unipolar endo heads surgical technique guide (stg). This +5mm offset should be taken into consideration when using a Cathcart head in a construct for hip arthroplasty procedures. Recall start date: June 7, 2022 |
Additional information
Details
Depuy Orthopaedics, Inc.
700 Orthopaedic Drive, Warsaw, Indiana, United States, 46582
Get notified