This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MOBILE DETECTOR HOLDER TROLLEY (2017-09-01)

Starting date:
September 1, 2017
Posting date:
September 26, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-64578

Affected products

MOBILE DETECTOR HOLDER TROLLEY

Reason

If the lock mechanism of the movable clamp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.

Affected products

MOBILE DETECTOR HOLDER TROLLEY

Lot or serial number

Not Applicable

Model or catalog number

Not Applicable

Companies
Manufacturer
Philips Medical Systems Dmc Gmbh
Rontgenstrabe 24
Hamburg
22335
GERMANY