MNA-240 ENC
Brand(s)
Last updated
Summary
Product
MNA-240 ENC
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| MNA-240 ENC | Not applicable | K9303078A K9303098A K9303097A |
Issue
The way of installing the MNA encoders caused the internal temperature in the encoder to exceed the maximum specification, which increased the risk of malfunction.
Recall start date: September 16, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
Barco N.V. Healthcare Division
President Kennedypark 35, Kortrijk, Belgium, 8500
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-72019
Get notified
Receive emails about new and updated recall and safety alerts.