MiniMed and Paradigm Insulin Pumps
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
MiniMed 670g |
All lots. |
MMT-1782K |
MiniMed 670g Kit |
All lots. |
MMT-1762KCN |
MiniMed 770g |
All lots. |
MMT-1881 |
MiniMed 630g |
All lots. |
MMT-1714K |
Paradigm 554 Insulin Pump |
All lots. |
MMT-554 |
MiniMed 780g Pump |
All lots. |
MMT-1885 |
MiniMed 780g Kit |
All lots. |
MMT-1895 |
Paradigm 754 Insulin Pump |
All lots. |
MMT-754 |
Issue
When air pressure decreases (e.g., during flight takeoff), more insulin may be released than expected. Additionally, unintended insulin may be released even if the pump's delivery is suspended or programmed to zero units per hour. When air pressure increases (e.g., during flight landing), less insulin may be released than expected.
Recall Start Date: January 28, 2025
Additional information
Details
Medtronic Minimed
18000 Devonshire St., Northridge, California, United States, 91325