MiniMed 670 g, 780 g, 770 g and 630 g
Brand(s)
Last updated
Summary
Product
MiniMed 670 g, 780 g, 770 g and 630 g
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Minimed 670g | All lots. | MMT-1782K MMT-1782 MMT-1782KRP MMT-1782CRP |
Minimed 670g Kit | All lots. | MMT-1762KCN MMT-1762CRP |
Minimed 780g Kit | All lots. | MMT-1895CN |
Minimed 630g | All lots. | MMT-1714K MMT-1714KRP MMT-1714 |
Minimed 770g | All lots. | MMT-1881 MMT-1891CN |
Issue
In some instances, pumps that have been dropped, bumped, or experienced physical impact, may experience damage to internal electrical components, which may result in reduced battery life, causing the battery alerts to occur when less battery life remains than the user guide states.
Recall start date: October 11, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Medtronic Minimed |
18000 Devonshire St., Northridge, California, United States, 91325 |
Published by
Health Canada
Audience
General public
Recall class
Type I
Identification number
RA-76339
Get notified