Health product recall

MiniMed 670 g, 780 g, 770 g and 630 g

Last updated

Summary

Product
MiniMed 670 g, 780 g, 770 g and 630 g
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products Lot or serial number Model or catalogue number
Minimed 670g All lots. MMT-1782K MMT-1782 MMT-1782KRP MMT-1782CRP
Minimed 670g Kit All lots. MMT-1762KCN MMT-1762CRP
Minimed 780g Kit All lots. MMT-1895CN
Minimed 630g All lots. MMT-1714K MMT-1714KRP MMT-1714
Minimed 770g All lots. MMT-1881 MMT-1891CN

Issue

In some instances, pumps that have been dropped, bumped, or experienced physical impact, may experience damage to internal electrical components, which may result in reduced battery life, causing the battery alerts to occur when less battery life remains than the user guide states.

Recall start date: October 11, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Medtronic Minimed
18000 Devonshire St., Northridge, California, United States, 91325
Published by
Health Canada
Audience
General public
Recall class
Type I
Identification number
RA-76339

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe