Minicap Extended Life Pd Transfer Set Clamps
Brand(s)
Last updated
Summary
Product
Minicap Extended Life Pd Transfer Set Clamps
Issue
Medical devices - Increased risk or new adverse events
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalog number |
|---|---|---|
| Minicap Extended Life Pd Transfer Set With Roller Clamp (15cm) | All lots. | 5C4449 |
| Minicap Extended Life Pd Transfer Set With Twist Clamp (Extra Short) (10cm) | All lots. | 5C4483 |
| Minicap Extended Life Pd Transfer Set With Twist Clamp (15cm) | All lots. | 5C4482 |
| Minicap Extended Life Pd Transfer Set With Twist Clamp (Extra Long) (22cm) | All lots. | R5C4484 |
Issue
Baxter Corporation is issuing an urgent medical device correction for the Minicap Extended Life Pd Transfer Sets. The following products may cause damage if they come into direct contact with the transfer set: Cleaning products such as hand sanitizer and solvents intended to remove adhesive residue.
Recall start date: Jan 12, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Companies
Baxter Healthcare Corporation
1 Baxter Parkway, Deerfield, Illinois
United States, 60015
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63857
Get notified
Receive emails about new and updated recall and safety alerts.