Health product recall

MIM (Software Package)

Last updated

Summary

Product
MIM (Software Package)
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
MIM (Software Package) N/A 7.2

Issue

In situations where two images with differing fields of view (FOV) complete an image fusion, an incorrect, elevated maximum standardized uptake value (SUV) could result. This could lead to mistaken assessment of disease status, progression or response to treatment and potential inappropriate therapy.

Recall Start Date: February 21, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

MIM Software Inc.

25800 Science Park Drive, Cleveland, Ohio, United States, 44122

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-77076

Get notified

Receive notifications for new and updated recalls and alerts by category.