Milling Handpiece (2016-10-06)
- Starting date:
- October 6, 2016
- Posting date:
- February 1, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65890
Affected products
Milling Handpiece
Reason
Zimmer Biomet is conducting a medical device field removal for all distributed milling handpieces. Complaints have been reported that the handpiece is inoperable. The investigation found that the complaints were related to the lack of preventative maintenance contrary to what is recommended in the instructions for use included with the handpiece.
Affected products
Milling Handpiece
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
00-5927-040-00