Health product recall

Millennium MLC 120 (2019-06-27)

Starting date:
June 27, 2019
Posting date:
August 30, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70837


Last updated:

Affected Products

Millennium MLC 120

Reason

After a recent upgrade to Multileaf Collimator [MLC] software version 8.5, A site reported that the MLC leaves did not move during an arc treatment.

Affected products

Millennium MLC 120

Lot or serial number

Serial number: H542967

Serial number: H544081

Serial number: H544352

Serial number: H544783

Serial number: H546262

Serial number: H546362

Serial number: H546677

Model or catalog number

MILLENNIUM MLC 120

Companies
Manufacturer

Varian Medical Systems, Inc.

3100 Hansen Way

Palo Alto

94304

California

UNITED STATES