Health product recall

Milford Smartbase (2018-03-27)

Starting date:
March 27, 2018
Posting date:
May 4, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-66682

Affected products

Milford Smartbase

Reason

Autochair's evaluation found that the combination of a weak spring and incorrect alignment of the locking pin has led to the legs closing fully when a side load was applied. This resulted in the device to become unstable and fall. The collapse of a Smartbase when in use, will lead to a user fall.

Affected products

Milford Smartbase

Lot or serial number

All lots.

Model or catalog number

9934105

Companies
Manufacturer
Autochair Ltd
Meadow Lane Industrial Estate
Alfreton, Derbyshire
DE55 7JR
UNITED KINGDOM