Health product recall

MIDAS REX MICROSAWS (2016-05-18)

Starting date:
May 18, 2016
Posting date:
June 30, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59110

Affected products

A. MIDAS REX MICROSAWS

Reason

Medtronic powered surgical solutions testing of the sagittal saw rocker arm seal identified a potential for leakage of fluid into the motor during use.  There have been no reports from the field of leakage into or out of the sagittal saw.  In assessing the risk of this potential failure mode, it was determined that the gravity sterilization cycle included in the IPC user manual may not be adequate to ensure sterility of the saws if fluid ingress occurs beyond the rocker housing seal.

Affected products

A. MIDAS REX MICROSAWS

Lot or serial number

P06274726

P06279710

Model or catalog number

ES300

ES310

Companies
Manufacturer

Medtronic Powered Surgical Solutions

4620 NORTH BEACH STREET

Fort Worth

76137

Texas

UNITED STATES