Health product recall

MICROSTREAM CO2 EXTENSION MODULE (2020-06-05)



Starting date:
June 5, 2020
Posting date:
September 25, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73985


Last updated:

Affected Products

MICROSTREAM CO2 EXTENSION MODULE

Reason

Affected M3015A extensions are incompatible with the IntelliVue X3 patient monitor (model no. 367030) and IntelliVue MMX Multi-measurement server. If the extension is connected to an X3 or MMX, which is in turn connected to a host monitor, communication between the X3/MMX and host monitor could be disrupted, resulting in a loss of monitoring and one of the following two technical alarms on the host monitor:

< X3 unplugged >

< MMX unplugged >

Affected products

MICROSTREAM CO2 EXTENSION MODULE

Lot or serial number

All lots.

Model or catalog number

M3015A

Companies
Manufacturer

Philips Medizin Systeme Boeblingen GMBH

HEWLETT-PACKARD STR. 2,

BOEBLINGEN

71034

GERMANY