MICROSTREAM CO2 EXTENSION MODULE (2020-06-05)
- Starting date:
- June 5, 2020
- Posting date:
- September 25, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-73985
Last updated:
2020-09-25
Affected Products
MICROSTREAM CO2 EXTENSION MODULE
Reason
Affected M3015A extensions are incompatible with the IntelliVue X3 patient monitor (model no. 367030) and IntelliVue MMX Multi-measurement server. If the extension is connected to an X3 or MMX, which is in turn connected to a host monitor, communication between the X3/MMX and host monitor could be disrupted, resulting in a loss of monitoring and one of the following two technical alarms on the host monitor:
< X3 unplugged >
< MMX unplugged >
Affected products
MICROSTREAM CO2 EXTENSION MODULE
Lot or serial number
All lots.
Model or catalog number
M3015A