This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MICROGAS 7650 BASIC UNIT (2015-07-03)

Starting date:
July 3, 2015
Posting date:
August 10, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54526

Affected Products

A. MICROGAS 7650 BASIC UNIT

Reason

It has recently been discovered that the storage temperature range for the Combi M Sensor 82 is stated incorrectly in the operating manual for the MicroGas 7650. The operation manual incorrectly states that the storage temperature range is -40°C to +70°C. The correct temperature range is -10°C to +50°C.

Affected products

A. MICROGAS 7650 BASIC UNIT

Lot or serial number

N/A

Model or catalog number
  • 0604210
  • 0605000
  • 0605200
Companies
Manufacturer
Radiometer Medical APS
AUSTRASSE 25 Basel
BASELLAND
SWITZERLAND