Health product recall

Microdeep Depth Electrode

Brand(s)
Last updated

Summary

Product
Microdeep Depth Electrode
Issue
Medical devices - Performance
What to do

Retrieve and isolate the affected products.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Microdeep Depth Electrode

More than 10 numbers, contact manufacturer.

D08-05AM

Microdeep Depth Electrode

More than 10 numbers, contact manufacturer.

D08-10AM

Microdeep Depth Electrode

More than 10 numbers, contact manufacturer.

D08-15CM

Microdeep Depth Electrode

More than 10 numbers, contact manufacturer.

D08-15AM

Issue

A manufacturing defect has been identified in the caps used in certain Dixi Medical devices. Although the cap is screwed onto the guiding screw, it is possible for the electrodes to slip inside. this indicates that the clamping function provided by the screw may not be fully effective, potentially compromising the safe positioning of the device. Customers are requested to isolate the affected devices.

Recall start date: July 30, 2025

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Neurology
Companies
Dixi Medical
2a Route De Pouligney Chaudefontaine, Marchaux-Chaudefontaine, France, 25640
Published by
Health Canada
Audience
Healthcare
Recall class
Type I
Identification number
RA-77884

Get notified

Receive emails about new and updated recall and safety alerts.