This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Micro SSP Allele Specific HLA Class I DNA Typing Tray-C*02 (2014-03-25)

Starting date:
March 25, 2014
Posting date:
August 28, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41233

Recalled products

  1. Micro SSP Allele Specific HLA Class I DNA Typing Tray-C*02 

Reason

Micro SSP Allele Specific HLA Class I DNA Typing Tray-C*02, the primer in well 1/7/c is incorrect.  As a result, the allele C*02:04 cannot be correctly assigned and may cause an ambiguous misassignment to C*02:27:01 and C*02:65.  These alleles are very rare occurring.

Affected products

A. Micro SSP Allele Specific HLA Class I DNA Typing Tray-C*02

Lot or serial number
  • 002, Batches 001-006
Model or catalog number
  • SSPR1-C02
Companies
Manufacturer

One Lambda Inc.

21001 Kittridge Street

Canoga Park

91303-2801

California