Health product recall

Michele Sleep Scoring System (2020-05-04)



Starting date:
May 4, 2020
Posting date:
May 29, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73203


Last updated:

Affected Products

Michele Sleep Scoring System

Reason

Vulnerabilities (CVE-2020-11651/CVE-2020-11652) identified in the software deployment tool used by the Michele Sleep

Scoring System allowed attackers to access and run arbitrary commands on both the 'master' and linux based 'minion' salt servers on 02- May-20. The attackers used this to stop the MICHELE software on these servers and run cryptocurrency mining software.

Affected products

Michele Sleep Scoring System

Lot or serial number

3.00.016

Model or catalog number

MICHELE

Companies
Manufacturer

Cerebra Medical

1470 Willson Place, Unit B

Winnipeg

R3T 3N9

Manitoba

CANADA