Mic-Peg Kit
Brand(s)
Last updated
Summary
Product
Mic-Peg Kit
Issue
Medical devices - Manufacturing defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
Mic-Peg Kit |
30135103 |
7640-20 |
Mic-Peg Kit |
30132546 |
8640-20 |
Mic-Peg Kit |
30132545 |
|
Mic-Peg Kit |
30132544 |
|
Mic-Peg Kit |
30132543 |
|
Mic-Peg Kit |
30129345 |
|
Mic-Peg Kit |
30116959 |
Issue
The External Retention Bolster contained in some of the mic* standard Percutaneous Endoscopic Gastrostomy Peg Kit - 20 FR - pull has a larger diameter (24FR) than the tube (20FR). As a result, the Bolster is too large to hold the Peg Tube to the correct tension which keeps it in place.
Recall start date: December 22nd, 2021.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices
Companies
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63789
Get notified
Receive notifications for new and updated recalls and alerts by category.