Health product recall

Miami J Select Cervical Collar

Brand(s)
Last updated

Summary

Product
Miami J Select Cervical Collar
Issue
Medical devices - Performance
What to do

Customers to be notified if they have affected devices, and what actions are to follow.

Audience
Healthcare
Industry

Affected products

Affected products Lot or serial number Model or catalogue number
Miami J Select Cervical Collar More than 10 numbers, contact manufacturer. MJS-101 MJSR-101

Issue

The PSA (pressure sensitive adhesive) hooks can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.

Recall start date: August 15, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
Ossur Hf
Grjothals 1-5, Reykjavik, Iceland, 110
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-75966

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe