Product
Miami J Select Cervical Collar
Issue
Medical devices - Performance
What to do

Customers to be notified if they have affected devices, and what actions are to follow.

Audience
Healthcare
Industry

Affected products

Affected products

Lot or serial number

Model or catalogue number

Miami J Select Cervical Collar

More than 10 numbers, contact manufacturer.

MJS-101
MJSR-101

Issue

The PSA (pressure sensitive adhesive) hooks can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.

Recall start date: August 15, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
Ossur Hf
Grjothals 1-5, Reykjavik, Iceland, 110
Published by
Health Canada
Audience
Healthcare
Industry
Recall class
Type II
Identification number
RA-75966