Mevatron, Oncor and Artiste Linear Accelerators (2017-01-31)
- Starting date:
- January 31, 2017
- Posting date:
- February 14, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Hospitals
- Identification number:
- RA-62206
Affected Products
- Mevatron Linear Accelerators
- Oncor Expression Digital Linear Accelerator
- Artiste MV-Linear Accelerator-Main Unit
Reason
It has come to our attention that for digital linear accelerators of type Artiste, Oncor and Primus with serial numbers equal or higher than 3094:
- Incorrect values for the rotational collimator position may be displayed due to mechanical problems of the sensor slide.
- The overhead suspension might break due to an improper weld seam, which might result in an injury of the patient.
Affected products
A. Mevatron Linear Accelerators
Lot or serial number
Serial numbers higher than or equal to 3094
Model or catalog number
4504200
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B. Oncor Expression Digital Linear Accelerator
Lot or serial number
Serial numbers higher than or equal to 3094
Model or catalog number
7360717
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
C. Artiste MV-Linear Accelerator-Main Unit
Lot or serial number
Serial numbers higher than or equal to 3094
Model or catalog number
8139789
Companies
- Manufacturer
-
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY