This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Mevatron, Oncor and Artiste Linear Accelerators (2017-01-31)

Starting date:
January 31, 2017
Posting date:
February 14, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals
Identification number:
RA-62206

Affected Products

  1. Mevatron Linear Accelerators
  2. Oncor Expression Digital Linear Accelerator
  3. Artiste MV-Linear Accelerator-Main Unit

Reason

It has come to our attention that for digital linear accelerators of type Artiste, Oncor and Primus with serial numbers equal or higher than 3094:

  1. Incorrect values for the rotational collimator position may be displayed due to mechanical problems of the sensor slide.
  2. The overhead suspension might break due to an improper weld seam, which might result in an injury of the patient.
 

Affected products

A. Mevatron Linear Accelerators

Lot or serial number

Serial numbers higher than or equal to 3094

Model or catalog number

4504200

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY



B. Oncor Expression Digital Linear Accelerator

Lot or serial number

Serial numbers higher than or equal to 3094

Model or catalog number

7360717

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY



C. Artiste MV-Linear Accelerator-Main Unit

Lot or serial number

Serial numbers higher than or equal to 3094

Model or catalog number

8139789

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY