This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Merge IConnect Enterprise Archive (ICEA)

Starting date:
August 26, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60242

Affected Products

Merge IConnect Enterprise Archive (ICEA)

Reason

This recall has been initiated because these versions of (ICEA) allow images to be stored to the ICEA without pixel data. A side effect of images without pixel data is performance degradation caused by the system continuously attempting to route the images to the peer system that fail to store.

Affected products

Merge IConnect Enterprise Archive (ICEA)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

40020

Companies
Manufacturer

Merge Healthcare Incorporated,

900 Walnut Ridge Drive,

Hartland

53029

Wisconsin

UNITED STATES