MERGE EYE STATION (2018-12-20)
- Starting date:
- December 20, 2018
- Posting date:
- August 9, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70655
Last updated: 2019-08-09
Affected Products
MERGE EYE STATION
Reason
Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
Affected products
MERGE EYE STATION
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
MERGE EYE STATION
Companies
- Manufacturer
-
Merge Healthcare Incorporated,
900 Walnut Ridge Drive,
Hartland
53029
Wisconsin
UNITED STATES