Health product recall

MERGE EYE STATION (2018-12-20)

Starting date:
December 20, 2018
Posting date:
August 9, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70655



Last updated: 2019-08-09

Affected Products

MERGE EYE STATION

Reason

Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).

Affected products

MERGE EYE STATION

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

MERGE EYE STATION

Companies
Manufacturer

Merge Healthcare Incorporated,

900 Walnut Ridge Drive,

Hartland

53029

Wisconsin

UNITED STATES