Health product recall

Megadyne Disposable Hand-Control Suction Coagulator With Cord And Holster

Last updated

Summary

Product
Megadyne Disposable Hand-Control Suction Coagulator With Cord And Holster
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Disposable Hand-Control Suction Coagulator With Cord And Holster All lots. 0043-25 0042-25 0041-25

Issue

Megadyne identified a potential issue during internal testing on the suction coagulator product family which may lead to fluid ingress into the handpiece. Fluid ingress can lead to intermittent device activation, non-activation, or self-activation when plugged into the electrosurgery unit.

Recall start date : June 14, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies

Megadyne Medical Products Inc

4545 Creek Road, Cincinnati, Ohio, United States, 45242

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-73873

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe