Mega Soft Universal Patient Return Electrode
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Mega Soft Universal Patient Return Electrode |
All lots. |
0840 |
Issue
Megadyne has received reports of patient burn injuries in procedures where the Mega Soft Patient Return Electrodes were used. A root cause investigation on the reports included testing which showed a combination of factors when present together may result in potential for thermal injuries. The combination of these conditions may be more likely when the pad is used with infants and small children.
Recall Start Date: May 8, 2024
Additional information
Details
Megadyne Medical Products Inc
4545 Creek Road, Cincinnati, Ohio, United States, 45242
Get notified
Receive emails about new and updated recall and safety alerts.