MEERA, YUNO & YUNO II (2020-09-11)
- Starting date:
- September 11, 2020
- Posting date:
- October 15, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74131
Last updated:
Affected Products
A. MEERA
B. YUNO
C. YUNO II
Reason
As part of our post market surveillance, cases with pulled-out plastic sockets (IEC 60320-1) of mobile OR-Tables MEERA, YUNO and YUNO II were reported to Getinge. Electric parts and / or wires were exposed.
Affected products
A. MEERA
Lot or serial number
Serial numbers 221 to 226 and 701 to 711.
Model or catalog number
720001F0
Companies
- Manufacturer
-
Maquet GmbH
Kehler Strasse 31
Rastatt
76437
GERMANY
B. YUNO
Lot or serial number
Serial numbers 1 to 90.
Model or catalog number
143301F0
Companies
- Manufacturer
-
Maquet GmbH
Kehler Strasse 31
Rastatt
76437
GERMANY
C. YUNO II
Lot or serial number
Serial numbers 1 to 13.
Model or catalog number
143302F0
Companies
- Manufacturer
-
Maquet GmbH
Kehler Strasse 31
Rastatt
76437
GERMANY