Meera Operating Tables
Brand(s)
Last updated
Summary
Product
Meera Operating Tables
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
Meera Operating Tables | More than 10 numbers, contact manufacturer. | 7200.01F2 |
Issue
Under certain conditions, it has been identified that an issue might prevent the device from performing as intended. a specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. of the Meera OR-Table.
Recall start date: August 16, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Physical medicine
Companies
Maquet GmbH
Kehler Strasse 31, Rastatt, Germany, 76437
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74210
Get notified
Receive notifications for new and updated recalls and alerts by category.