Health product recall

Meera Operating Tables (2019-04-04)

Starting date:
April 4, 2019
Posting date:
May 3, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-69824


Last updated:

Affected Products

Meera Operating Tables

Reason

As part of post market surveillance, cases were reported to Getinge involving a mobile OR-Table Meera, and in which the plastic socket of the IEC 60320-1 socket was pulled out of the metal cage and electronic parts were exposed. In cases where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components.

Affected products

Meera Operating Tables

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

720001F0

720001F2

Companies
Manufacturer

Maquet GmbH

Kehler Strasse 31

Rastatt