Health product recall

MEDTRACT ST65 (2019-03-14)

Starting date:
March 14, 2019
Posting date:
May 24, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70019



Last updated: 2019-05-24

Affected Products

MEDTRACT ST65

Reason

An alternate component for an IGBT (Insulated Gate Bipolar Transistor) was added to a Printed Circuit Board Assembly (PCBA) of some treadmills manufactured between 22-dec-2017 and 03-dec-2018. In some instances, the alternate component will overheat, causing the treadmill to no longer function and potentially stop during a stress test. When the failure occurs while performing a test, the treadmill decelerates gradually following normal protocol.

Affected products

MEDTRACT ST65

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

TM55HEXRCE

TM55LEXRCE

Companies
Manufacturer

Mortara Instrument Inc.,

7865 North 86th St.,

Milwaukee

53224

Wisconsin

UNITED STATES