MEDTRACT ST65 (2019-03-14)
- Starting date:
- March 14, 2019
- Posting date:
- May 24, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70019
Last updated: 2019-05-24
Affected Products
MEDTRACT ST65
Reason
An alternate component for an IGBT (Insulated Gate Bipolar Transistor) was added to a Printed Circuit Board Assembly (PCBA) of some treadmills manufactured between 22-dec-2017 and 03-dec-2018. In some instances, the alternate component will overheat, causing the treadmill to no longer function and potentially stop during a stress test. When the failure occurs while performing a test, the treadmill decelerates gradually following normal protocol.
Affected products
MEDTRACT ST65
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
TM55HEXRCE
TM55LEXRCE
Companies
- Manufacturer
-
Mortara Instrument Inc.,
7865 North 86th St.,
Milwaukee
53224
Wisconsin
UNITED STATES