Medpor Titan Orbital Floor And Wall (Mtb)
Brand(s)
Last updated
Summary
Product
Medpor Titan Orbital Floor And Wall (Mtb)
Issue
Medical devices - Unauthorised device
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected Products | Lot or serial number | Model or catalog number |
|---|---|---|
| Medpor Titan Orbital Floor And Wall (Mtb) | 3R7YHD | 81032 |
Issue
|
On November 22, 2021 via customer complaint - Stryker became aware the inside the packaging of one reported item #81032 (Titan OFW-MTB, right) a non-conforming product was found. Contrary to the design, the barrier layer was manufactured on the inferior side of the implant versus the superior side. Recall start date: 2021-12-21 |
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General and plastic surgery
Companies
| Stryker Leibinger Gmbh & Co. Kg |
| Botzinger Strasse 41, Freiburg, Germany, 79111 |