Medline ENFit G-Tube Connector
Brand(s)
Last updated
Summary
Product
Medline ENFit G-Tube Connector
Issue
Medical devices - Device compatibility
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
ENFit G-Tube Connector |
All lots. |
ENFIT1010GC |
Issue
Medline ENFit G-Tube Connector (ENFIT1010GC) is recalled. This connector was not manufactured to the required dimensional specifications. Thus, the connector may not form a proper seal with ENFit style devices, so lead to leakage. This may delay the patient care, particularly during medication delivery. Customers should destroy this product.
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies
Medline Industries, LP
Three Lakes Drive, Northfield, Illinois, United States, 60093
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81862
Get notified
Receive emails about new and updated recall and safety alerts.