This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

MEDLINE CUSTOM UROLOGY PROCEDURE KIT/ TRAY (2016-08-11)

Starting date:
August 11, 2016
Posting date:
August 19, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59900

Affected products

  1. TURP PACK BGH-LF
  2. TURP SJHC-LF
  3. TUR PMH-LF

Reason

The Medline packs contain bard irrigation syringes under recall Bard.  Bard is recalling the syringes because the syringe may be at risk for a slit defect in the packaging that may affect sterility. The Bard syringe in these packs is to be removed and discarded. The Medline packs are terminally sterilized, therefore the remaining components within the packs are safe to use.

Affected products

  1. TURP PACK BGH-LF

Lot or serial number

15HB7157

15VB3976

Model or catalog number

DYNJ41729

Companies
Manufacturer

Medline Industries Inc

1 Medline Place

Mundelein

Illinois

UNITED STATES



B. TURP SJHC-LF

Lot or serial number

15GB9759

15UB2552

15WB8225

Model or catalog number

DYNJ46842

DYNJ46842A

Companies
Manufacturer

Medline Industries Inc

1 Medline Place

Mundelein

Illinois

UNITED STATES



C. TUR PMH-LF

Lot or serial number

15VB4469

15XB6975

Model or catalog number

DYNJ900672

Companies
Manufacturer

Medline Industries Inc

1 Medline Place

Mundelein

Illinois

UNITED STATES