Health product recall

Medline Convenience Kits Containing Cannula Components with Female Luer Lock Connectors

Last updated

Summary

Product
Medline Convenience Kits Containing Cannula Components with Female Luer Lock Connectors
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information.

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

EM Pop Pack

25JBB049

DYNJ52921B

Pontage-LF

More than 10 numbers, contact manufacturer.

DYNJ58124D

Pontage-LF

More than 10 numbers, contact manufacturer.

DYNJ58124B

Pontage-LF

More than 10 numbers, contact manufacturer.

DYNJ58124C

Issue

Specific item(s) and lot(s) of Medline Convenience Kits Containing Cannula Components with Female Luer Lock Connectors are being recalled. Medline has identified that the components do not conform to ISO 80369-7 dimensional requirements. The nonconforming dimensions may result in a loose, unstable, or detached Luer Lock connection and, in certain cases, may contribute to leaking or breaking during use, so customers will be directed to remove and destroy affected components.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Medline Industries, Lp

Three Lakes Drive, Northfield, Illinois, United States, 60093

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-82712