Medihoney® Antibacterial Dressing
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Medihoney® Antibacterial Dressing |
All lots. |
391 |
Medihoney® Antibacterial Dressing |
All lots. |
395 |
Medihoney® Antibacterial Dressing |
All lots. |
799 |
Medihoney® Antibacterial Dressing |
All lots. |
784 |
Medihoney® Antibacterial Dressing |
All lots. |
781 |
Medihoney® Antibacterial Dressing |
All lots. |
780 |
Medihoney® Antibacterial Dressing |
All lots. |
785 |
Medihoney® Antibacterial Dressing |
All lots. |
398 |
Medihoney® Antibacterial Dressing |
All lots. |
782 |
Medihoney® Antibacterial Dressing |
All lots. |
787 |
Medihoney® Antibacterial Dressing |
All lots. |
783 |
Issue
This voluntary recall is due to the potential for breaches to the sterile barrier identified during packaging testing, which creates a potential sterility concern. The specific potential failures are one or several of the following:
- inadequate sealing of sterile barrier packaging
- shipping box does not adequately protect device during transportation
- tube twist-off cap failure
Recall start date: August 14, 2025
Additional information
Details
Derma Sciences, Inc.
104 Shorting Road, Toronto, Ontario, Canada, M1S 3S4
Get notified
Receive emails about new and updated recall and safety alerts.