Health product recall

Med-RX® I.V. Set Stopcock

Last updated

Summary

Product
Med-RX® I.V. Set Stopcock
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number
Med-RX® I.V. Set Stopcock 176187 184143 176857 10-8710

Issue

Smith Medical has issuing a recall of its Logical 3 Gang Manifold due to potential defect with the rotating   adaptor(rotator) which is incorporated as a component within the product. It was identified that the inner diameter of the device's o-ring may be oversized affecting seal integrity.  The recalled product was used in the manufacture of one of our device named above.

Recall start date: November 21, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - General hospital and personal use
Companies

Canadian Hospital Specialties Ltd. Doing Business As CHS

2760 Brighton Road, Oakville, Ontario, Canada, L6H 5T4

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74716

Get notified

Receive emails about new and updated recall and safety alerts.