MED-RX® Epidural Anesthesia Kit
Last updated
Summary
Product
MED-RX® Epidural Anesthesia Kit
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
MED-RX® Epidural Anesthesia Kit |
191306 |
80-8169 |
Issue
The depth markers on the epidural catheter have potential to fade or wear away during product use.
This issue could create a delay in procedure, as it may prevent anesthesiologists from accurately determining the depth and placement of the catheter during the anesthesia procedure which may require product replacement.
Recall start date: April 30, 2024
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Companies
Canadian Hospital Specialties Ltd. Doing Business As CHS
2760 Brighton Road, Oakville, Ontario, Canada, L6H 5T4
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75503
Get notified