Health product recall

Med-RX Wound Care Kit (2020-01-07)



Starting date:
January 7, 2020
Posting date:
February 14, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72313


Last updated:

Affected Products

Med-RX Wound Care Kit

Reason

Notification  due to a component manufacturer recall on the one-link needle-less connector where there is concern that more than expected force is required to create a secure connection and avoid unintentional disconnections. Insufficient connection and tightening force may lead to disconnection. Manufacturer has noted that when the one-link connectors are swabbed as instructed in the IFU, the applied antiseptic reduces the force required to achieve a secure connection.

Affected products

Med-RX Wound Care Kit

Lot or serial number

All lots.

Model or catalog number

80-7093

Companies
Manufacturer

Canadian Hospital Specialties Ltd. doing business as CHS

2760 Brighton Road

Oakville

L6H 5T4

Ontario

CANADA