Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41337

Recalled Products

Mckesson Cardiology Hemo 

Reason

Incorrect hemoglobin values for hemo calculations.  Mckesson has identified a scenario in which the Mckesson Cardiology Hemo calculation section incorrectly converts the hemoglobin value before it is utilized in the applicable formula calculations.  As a result, the formula output is incorrect.

Affected products

Mckesson Cardiology Hemo 

Lot or serial number

MC13.0 HF3

Model or catalog number

13.0
13.0HF1
13.0HF2
13.0HF3
13.1