Mckesson Cardiology Hemo (2014-08-27)
- Starting date:
- August 27, 2014
- Posting date:
- September 11, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-41337
Recalled Products
Mckesson Cardiology Hemo
Reason
Incorrect hemoglobin values for hemo calculations. Mckesson has identified a scenario in which the Mckesson Cardiology Hemo calculation section incorrectly converts the hemoglobin value before it is utilized in the applicable formula calculations. As a result, the formula output is incorrect.
Affected products
Mckesson Cardiology Hemo
Lot or serial number
MC13.0 HF3
Model or catalog number
13.0
13.0HF1
13.0HF2
13.0HF3
13.1
Companies
- Manufacturer
-
Mckesson Medical Imaging Company
130 - 10711 Cambie Road
Richmond
V6X 3G5
British Columbia
CANADA