McGRATH™ Mac 2 Laryngoscope Blade
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products | Lot or serial number | Model or catalog number |
---|---|---|
McGRATH™ Mac 2 Laryngoscope Blade | 22102001 22082207 22083101 | 350-017-000 350-084-000 |
Issue
Medtronic received six (6) complaints of poor picture quality and foggy and/or blurred images on the McGRATH™ Mac 2 video laryngoscope screen during use of these blades. The reports indicated a delay to treatment while an alternate device was located and/or oral tissue injury due to multiple intubation attempts. Medtronic investigation of these customer reports identified a manufacturing issue which resulted in an ineffective application of the antifog agent on the blades. An ineffective application of the anti-fog coating can result in condensation accumulating on the blade during use, resulting in the image on the video laryngoscope screen being blurred or obscured.
Recall start date: August 8, 2023
Additional information
Details
Covidien LLC
15 Hampshire Street, Mansfield, Massachusetts, United States, 02048